11 / 27 page ![]() TEK.COM 11 DMM 7510 7½-Digit Graphical Sampling Multimeter Accuracy Range 38 Resolution Maximum Burden Voltage Accuracy ±(ppm of reading + ppm of range) 24 Hour T CAL ±1°C 39 90 Day T CAL ±5°C 1 Year T CAL ±5°C 2 Year T CAL ±5°C Temperature Coefficient 40 10.000000 µA 1 pA 15 mV 30 + 30 100 + 30 125 + 40 175 + 50 10 + 8 100.00000 µA 10 pA 15 mV 20 + 5 75 + 12 100 + 15 150 + 20 10 + 3 1.0000000 mA 100 pA 15 mV 30 + 5 75 + 12 100 + 15 150 + 20 10 + 3 10.000000 mA 1 nA 20 mV 40 + 5 75 + 12 100 + 15 150 + 20 10 + 3 100.00000 mA 10 nA 200 mV 50 + 18 300 + 30 400 + 30 500 + 30 50 + 5 1.0000000 A 100 nA 400 mV 150 + 50 400 + 50 450 + 50 500 + 50 10 + 10 3.000000 A 1 µA 1300 mV 200 + 40 400 + 40 450 + 40 500 + 40 10 + 10 10.000000 A 41 1 µA 650 mV 700 + 275 800 + 275 1500 + 275 2000 + 275 50 + 10 RMS Noise (additional peak noise uncertainty) 42 Applies to ± ppm of range. Peak noise uncertainty is included in DC specifications for ≥1 PLC. Add peak noise uncertainty to measurements when <1 PLC. Examples 1 mA at 0.006 PLC: 9 (from Accuracy table) + 20 (additional peak noise uncertainty) = 29 ppm of range. 1 mA at 1 PLC: 9 +0 = 9 ppm of range. NPLC Digits 10 µA 100 µA 1 mA 10 mA 100 mA 1A 3A 10A 43 5 7½ 0.15 0.14 0.09 0.1 0.3 0.3 0.2 0.8 1 7½ 0.4 0.13 0.1 0.1 0.5 0.5 0.3 1.2 0.2 6½ 0 (220) 0 (23) 0.2 (3.4) 0.2 (1.6) 2 (10) 2 (11) 0.7 (4.6) 4 (32) 0.2 44 6½ 120 (260) 12 (26) 1.2 (3.8) 0.3 (1.8) 1.9 (9.8) 2 (10) 0.8 (5) 8 (37) 0.06 5½ 130 (280) 12 (29) 1.3 (5.6) 0.4 (3.9) 2 (14) 2 (14) 1.2 (7.7) 10 (59) 0.006 4½ 130 (350) 14 (42) 3 (20) 2 (20) 4 (30) 4 (31) 7 (51) 20 (110) 0.0005 3½ 260 (2110) 30 (300) 20 (150) 20 (160) 30 (190) 30 (190) 70 (510) 60 (420) DC Current Characteristics Range 10 µA 100 µA 1 mA 10 mA 100 mA 1 A 3 A 10 A 43 Effective Internal Shunt Value 45 1 k Ω 100 Ω 10 Ω 1 Ω 0.1 Ω 0.1 Ω 0.1 Ω 0.005 Ω Autozero Off Error: For ±1°C and ≤10 minutes add ±(8 ppm of reading + range error) 150 pA 1.5 nA 15 nA 150 nA 15 µA 150 µA 150 µA 3 mA Overload Recovery: For each additional sustained amp beyond ±1.5 A, add the following initial ppm of range error until thermally settled after overload recovery 15500 1800 150 150 6500 200 — — NOTES 42. Noise values are based on 1000 readings with autozero on and AMPS terminal open. RMS noise is typical. Additional peak noise is guaranteed. 43. Rear input terminals only. 44. With line sync on. 45. Values are typical and guaranteed by design. |
|
Ligação URL |
| ALLDATASHEET é útil para você? [ DONATE ] |
Sobre Alldatasheet | Publicidade | Contato conosco | Privacy Policy | Link para a ficha técnica | roca de Link | Lista de Fabricantes All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |